8-5. Calibration Test Using An External Weight

This function tests the weighing accuracy using an external
weight and outputs the result. This is available only when
the "GLP output (

info

)" parameter is set to "

1

" or "

2

".

(Calibration test does not perform calibration)

Step 1 Connect the AC adapter and warm up the balance

for at least 30 minutes with nothing on the weighing
pan.

Step 2 Press and hold the

CAL

key until

CCout

is

displayed, then release the key.

Step 3 The balance displays

CC 0 .

If you want to change the target weight, press

the

SAMPLE

key and go to step 4.

A list of usable weights is shown on page 25.

If you use the target weight value stored in

the balance, go to step 5.

Step 4 Specify the target weight value as follows:

SAMPLE

key... To switch the display condition to: "All of the

segments blinking" (target weight selection
mode) or "The last two digits blinking" (value
adjustment mode).

RE-ZERO

key . To select the target weight or adjust the value.

In the value adjustment mode, -15 digits
appear after +15 digits.

PRINT

key...... To store the new weight value. Even if the AC adapter is

removed, the data is maintained in non-volatile memory.

CAL

key ......... To cancel the operation and return to

CC 0

.

Step 5 Confirm that there is nothing on the pan and press

the

PRINT

key. The balance measures the zero

point and displays the measured value. Do not
allow vibration or drafts to affect the balance. The
balance displays the target weight value.

Step 6 Place the displayed target weight on the pan and

press the

PRINT

key. The balance measures the

target weight and displays the measured value. Do
not allow vibration or drafts to affect the balance.

Step 7 The balance displays

end

.

Remove the weight from the pan.

Step 8 The balance displays

glp

and outputs "calibration

test report" using the RS-232C interface or stores the
calibration test data in memory. Refer to "11-2 GLP
Report" of the function table for details.

Step 9 The balance will automatically return to the weighing mode.

Release

Press and hold

Select

Place weight

Remove

GLP output

Example:

New weight

10001.2 g