12. Data Memory

Data memory is a function to store weighing data, calibration data and unit mass in memory.
Of the data in memory, the balance can only display the weighing data. The weighing data
and calibration data in memory are available for outputting at one time to a printer or
personal computer. One of the following data sets can be stored:

GX-K series

GF-K series

Excluding date and time

Up to 200 sets Up to 40 sets

Weighing data

Including date and time

Up to 100 sets

-

Calibration result of Internal and external calibration
and calibration test

Last 50 sets

-

Unit mass in the counting mode

Up to 50 sets Up to 20 sets

Upper and lower limit values of comparator function

Up to 20 sets

Tare value

Up to 20 sets


Note GF-K series does not output time and date.

12-1. Notes on Using Data Memory

To use the memory function, set the "Data memory (

data

)" parameter of the function table.

In addition, for weighing data, set the "Time/Date output (

5-td

)" parameter. For details on

setting the data memory, refer to "10-2. The Function Table".

For weighing data of the GX-K series, the data contents to be stored and the storage

capacity depend on the "Time/Date output (

5-td

)" parameter setting.

Releasing "Clr"

If a different type of data exists in memory when the data is

stored, "

Clr

" blinks the upper left of the display. For example,

you want to store weighing data but calibration data or unit mass
data remains in memory.

Under such a condition, before storing data, delete the data in memory as follows:

Step 1

Press and hold the

PRINT

key until

SClr "n"o

with "no" blinking is displayed,

then release the key.

Step 2 Press the

RE-ZERO

key to display

SClr "g"o

with "go" blinking.

The type of data stored in memory appears in the upper left of the display as

shown below:

Unit mass in the counting mode

PpC

Weighing data without time and date

-d-

Weighing data with time and date

d-t

Calibration result

Hi5

parameters of comparator

Cp

Tare value

tr

Step 3 Press the

PRINT

key to delete all the data in memory.

Step 4 The balance displays

end

and returns to the weighing mode.