3. after use, 4. power supply – A&D Weighing GF-32K User Manual

Page 11

Advertising
background image

9

Prevent foreign matter, such as powder, liquid and metal, from invading the area around

the weighing pan.

Use the "breeze break" for a precision weighing.


3-3. After Use

Avoid mechanical shock to the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service

or repair.

Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that

is moistened with warm water and a mild detergent.

Do not allow the balance to be immersed in water. Even though the balance complies with

IP code, the balance will not withstand being completely immersed in water.

The weighing pan, pan support and draft gate can be removed to clean the balance. Clean

by splashing with water.

Use the waterproof option RS-232C cable GX-07K, when RS-232C interface is used with

IP-65. AX-KO1710-200, a standard RS-232C cable, is not waterproof or dustproof.


3-4. Power Supply

Do not remove the AC adapter while the internal mass is in motion, for example, right

after the AC adapter is connected, or during calibration using the internal mass.
If the AC adapter is removed under the conditions described above, the internal mass
will be left unsecured, that may cause mechanical damage when the balance is moved.
Before removing the AC adapter, press the

ON:OFF

key and confirm that zero is

displayed.

When the AC adapter is connected, the balance is in the standby mode if the standby

indicator is on. This is a normal state and does not harm the balance. For accurate
weighing, keep the AC adapter connected to the balance and AC power unless the
balance is not to be used for a long period of time.

Advertising
This manual is related to the following products: