5-1. storing the upper and lower limit values – A&D Weighing GF-32K User Manual

Page 66

Advertising
background image

64

12-5. Data Memory for Comparator Settings

The data memory function can store 20 sets of upper and lower limit values for the

comparator mode. The reference value or tolerance value for the comparator mode can not
be stored in memory.

The upper and lower limit values in memory can be recalled easily using the

MODE

key

and used for weighing.

The upper and lower limit values in memory can be recalled and changed.


Note
The recalled upper and lower limit values can be changed, using the digital input

mode or the weighing input mode. The digital input mode enters the upper and lower
limit values using the keys. The weighing input mode uses a sample to store the
upper and lower limit values.


12-5-1. Storing the upper and lower limit values

To store new upper and lower limit values: Recall the stored upper and lower limit values to
be changed ("

C01

" to "

C20

") . Then, change the recalled upper and lower limit values using

the digital input mode or the weighing input mode, and store the new value.


Note

While the data memory function is in use, unit selection using the

MODE

key is

not available.

If "

Clr

" appears blinking in the upper left of the display, delete the data in memory.

Refer to "12-1. Notes on Using Data Memory".

Step 1 Press the

MODE

key to select a unit to be

used for storage.

Step 2 Set the "Data memory (

data

)" parameter to "

4

".

Step 3 Press and hold the

PRINT

key until the

balance enters the upper and lower limit
values confirmation mode. The upper limit
value last selected is displayed.

Step 4 Select the comparator number to be used, using the following keys.

RE-ZERO

key..... To increase the comparator number by one.

MODE

key ......... To decrease the comparator number by one.

Each time the

RE-ZERO

key or

MODE

key is pressed, the displayed value

changes as follows:

C03

HI

C03

LO

C04

HI

C04

LO

Step 5

To change the selected upper and lower limit values:

To use the digital input mode, press the

SAMPLE

key to enter the digital input mode.

Go to "Digital input mode" on the next page.

To use the weighing input mode, press the

SAMPLE

key, then press and hold the

MODE

key to enter the weighing input mode. Go to "Weighing input mode" on the

next page.


No.3. Comparator number

Upper limit value

Advertising
This manual is related to the following products: