A&D Weighing GF-32K User Manual

Page 60

Advertising
background image

58

"

fUl

" indicates that memory is full or the memory capacity has been reached. More data can

not be stored unless the memory data is deleted.

Automatic self calibration can not be used while the interval memory mode is active.
The following commands can not be used during data storage.

Q

Query command for weighing data.

S

Query command for stable weighing data.

SI

Query command for weighing data.

SIR

Query command for continuous weighing data.


Setting the function table

Parameter settings for each output mode are as follows:

Item

Mode

Data output

mode

Auto print polarity,

difference

Data memory

function

Interval time

Key mode

prt 0

Not used

data 2

Auto print mode A

prt 1

data 2

Auto print mode B

prt 2

ap-a 0-2
ap-b 0-2

data 2

Not used

Interval memory mode

prt 3

Not used

data 2

int 0-8


Set each item for GX-K series, depending on the situation, as follows:

No

d-no 0

No

5-td 0

Data number

Yes

d-no 1

Time only

5-td 1

No

5-id 0

Date only

5-td 2

ID number

Yes

5-id 1

Time and date

Both

5-td 3

When 5-td 1, 2 or 3 is selected, the amount of data to be stored is 100 sets.

Set each item for GF-K series, depending on the situation, as follows:

No

d-no 0

Data number

Yes

d-no 1

No

5-id 0

ID number

Yes

5-id 1

GF-K series does not use time and date.

Enabling the data memory function

Step 1

Press and hold the

SAMPLE

key until

ba5fnc

is displayed, then release the key.

Step 2 Press

the

SAMPLE

key several times to display

dout

.

Step 3 Press

the

PRINT

key.

Step 4 Press

the

SAMPLE

key three times to display

data 0

.

Step 5 Press

the

RE-ZERO

key to display

data 2

.

Step 6 Press

the

PRINT

key to store the setting.

Step 7 Press

the

CAL

key to return to the weighing mode.

Advertising
This manual is related to the following products: