3. other display, 4. asking for repair – A&D Weighing GF-32K User Manual

Page 92

Advertising
background image

90

Display

Error code Description

EC,E03

Timeout error
If the timeout parameter is set to "

t-Up1

", the balance did not

receive the next character of a command within the time limit
of one second. Confirm the communication.

EC,E04

Excess characters error
The balance received excessive characters in a command.
Confirm the command.

EC,E06

Format error
A command includes incorrect data.
Example:

The data is numerically incorrect.

Confirm the command.

EC,E07

Parameter setting error
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.

Other error code

If an error described above can not be released or other
errors are displayed, contact the local A&D dealer.




19-3. Other Display

When this indicator blinks, automatic self calibration is required. The
indicator blinks when the balance detects changes in ambient
temperature. If the balance is not used for several minutes with this
indicator blinking, the balance performs automatic self calibration. The
blinking period depends on the operating environment.

Advise The balance can be used while this indicator is blinking. We

recommend that you perform automatic self calibration for precision
weighing.



19-4. Asking For Repair

If the balance needs service or repair, contact your local A&D dealer.

The balance is a precision instrument. Use much care when handling the balance and

observe the following when transporting the balance.

Use the original packing material for transportation.
Remove the draft gate, weighing pan and pan support from the main unit before placing the

balance in the shipping container.

Advertising