Specifications – A&D Weighing GF-32K User Manual

Page 93

Advertising
background image

91

20. Specifications

GX-8K GX-8K2 GX-10K GX-12K GX-20K GX-30K GX-32K

Weighing capacity

8.1 kg

10.1 kg

12 kg

21 kg

31 kg

31 kg

8.1008 kg

31.008 kg

Maximum display

8.10084 kg

2.10009 kg *1

10.10084 kg 12.0084 kg 21.0084 kg 31.0084 kg

6.1009 kg *1

Minimum weighing value (1 digit) 0.01 g 0.1 g / 0.01 g

0.01 g

0.1 g

1g / 0.1 g

Repeatability

(Standard deviation)

0.01 g

0.05 g /

0.01g

0.01 g

0.1 g

0.5g / 0.1 g

Linearity

±0.03 g

±0.1 g /

±0.02 g

±0.03 g

±0.2 g

±1g / ±0.2 g

Stabilization time

(Typical

at

FAST

)

Approx. 1.5 seconds

Sensitivity drift,

*2

(10°C ~ 30°C / 50

°F ~ 86°F) ±2 ppm/°C ±3 ppm/°C ±2 ppm/°C

±3 ppm/°C

±5 ppm/°C

Accuracy right after

calibration using the

internal mass

(Accuracy of full scale) *3

±0.15 g

±0.3 g

±0.15 g

±1.0 g

±1.5 g

±3 g

Operating environment

5°C to 40°C (41°F to 104°F), 85%RH or less (No condensation)

Internal mass

Built-in function

Weighing data of data

memory

200 data, 100 data with calendar

Time and clock function

Built-in function

Display refresh rate

5 times/second or 10 times/second

Minimum unit

mass

0.01 g

0.1 g

0.01 g

0.1 g

1g

Counting

mode Number of

samples

10, 25, 50 or 100 pieces

Minimum 100%

reference mass

1 g

10 g

1 g

10 g

100g

Percent mode

Minimum 100%

display

0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)

Interface

(Provided as standard)

RS-232C with Windows Communication Tools Software WinCT

External calibration weight

2 kg, 3 kg, 4 kg, 5 kg,

6 kg, 7 kg, 8 kg

2 kg, 3 kg,

4 kg, 5 kg,

6 kg, 7 kg,

8 kg, 9 kg,

10 kg

5 kg, 10 kg 10 kg, 20 kg

20 kg, 30 kg

Weighing pan

270 x 210 mm

External dimensions

300(W) x 355(D) x 111(H) mm

Power supply &

AC adapter type

Power consumption: Approx. 11VA (supplied to the AC adapter )

Confirm that the adapter type is correct for the local voltage and power receptacle type.

Weight

Approx. 9.3kg

Dust and water protection

Complying with IP65

*1: The balance allows weighing using a precision range, even with a heavy tare placed on the pan. (Smart range function)

*2: When automatic self calibration is not used.

*3: Accuracy right after calibration using the internal mass in good ambient conditions (within the temperature range of

10

°C to 30°C (50°F to 86°F) with no abrupt changes in temperature or humidity, no drafts, no effect by magnetic fields

or static electricity).

Check the internal mass periodically as described in "18. Maintenance".