11. ID Number And GLP Report

The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
The GLP output format is selected at "GLP output (

info

)" of the function table and can be

output to a personal computer or printer using the RS-232C serial interface.

The GLP output format includes the balance manufacturer, model, serial number, ID

number, date, time and space for signature for weighing data, the weight used and results
for calibration or calibration test data.

The balance can output the following reports for GLP.

"Calibration report" of the calibration, using the internal mass (Calibration due to changes

in temperature and one-touch calibration.)

"Calibration report" of the calibration, using an external weight.

"Calibration test report" of the calibration test, using an external weight.

"Title block" and "End block" for the weighing data.

Calibration and calibration test data can be stored in memory to output several reports at the

same time. Refer to "12. Data Memory" for details.

For details on confirming and setting the time and date for the GX-K series. Refer to "10-7.

Clock and Calendar Function".


Caution

GF-K series does not store the calibration report in memory.
GF-K series does not use the time and date function. Use the calendar function

of AD-8121B printer.


11-1. Setting The ID Number

Step 1 Press and hold the

SAMPLE

key until

ba5fnc

of the function table is displayed,

then release the key.

Step 2 Press the

SAMPLE

key several times to display

id

.

Step 3 Press the

PRINT

key. Set the ID number using the following keys.

RE-ZERO

key..... To set the character of the digit selected.

Refer to the display character set shown below.

SAMPLE

key ...... To select the digit to change the value.

PRINT

key ......... To store the new ID number and display

ba5fnc

.

CAL

key ............ To cancel the new ID number and display

ba5fnc

.


Step 4 With

ba5fnc

displayed, press the

CAL

key to return to the weighing mode.


Display character set