TROUBLESHOOTING & REPAIR

F-42

F-42

POWER WAVE 455M/MSTT

Retur

n to Section TOC

Retur

n to Section TOC

Retur

n to Section TOC

Retur

n to Section TOC

Retur

n to Master TOC

Retur

n to Master TOC

Retur

n to Master TOC

Retur

n to Master TOC

TABLE F.5 - CURRENT FEEDBACK AT VARIOUS OUTPUT LOADS

POWER WAVE CURRENT TRANSDUCER TEST (CONTINUED)

8. If for any reason the machine cannot be

loaded to 250 amps, Table F.5 shows what
feedback voltage is produced at various cur-
rent loads.

9. If the correct supply voltages are applied to

the current transducer, and with the machine
loaded, the feedback voltage is missing or

not correct, the current transducer may be
faulty. Also make certain that lead 211
(plug J8 pin 1) has continuity (zero ohms)
between the current transducer and the
control board. See the Wiring Diagram.

10. Install the right side case cover using the

3/8” nut driver.

OUTPUT LOAD CURRENT

500
450
400
350
300
250
200
150
100

50

EXPECTED TRANSDUCER FEEDBACK

VOLTAGE

4.0
3.6
3.2
2.8
2.4
2.0
1.6
1.2
0.8
0.4