Calibration

4-26

1760 Series Component Waveform/Vector Monitor User Manual

Set Horizontal Gain

1.

To set horizontal gain, push the

CONFIG

menu button.

Select

CALIBRATE

, then

CAL SIG ON

and

100 KHZ

.

Select

GAIN CAL ON

.

a.

In

1 LINE SWEEP

(5

µs/div), adjust the

HORIZ CAL

control for one-half cycle per major graticule
division; in

2 LINE SWEEP

(10

µs/div), adjust for one

full cycle per major division.

b.

Pushing

RESET H CAL

will cancel any gain adjust-

ment and reinstate the factory gain setting.

Set Vector Gain

1.

To adjust vector gain, select

VECTOR

display mode and

push the

CONFIG

menu button. Select

CALIBRATE

, then

CAL SIG ON

and

GAIN CAL ON

.

a.

Adjust

VECTOR CAL

so that the calibration signal

circle falls on the compass rose (outer ring).

b.

Pushing

RESET V CAL

will cancel any gain adjust-

ment and reinstate the factory gain setting.