Operator’s Checkout Procedure

1760 Series Component Waveform/Vector Monitor User Manual

2-15

4.

Check Gain

The 1760--Series internal calibrator can be used to check
instrument calibration. Select

CH-A

input and set the genera-

tor for NTSC Color Bars output (COMPOSITE for PAL).

Enter the configuration menu and select

CALIBRATE

. Turn on

the calibrator signal and note that the amplitude is 140 IRE
(1.0 V PAL), with one cycle per division. See Figure 2-6.

Select

VECTOR

display mode. Note that the calibrator signal

overlays the compass rose. Return to

WAVEFORM

display

mode.

Exit the configuration menu. Note that the calibrator signal
is automatically turned off. (For more information on using
the calibrator signal, see page 4-25.)