Using the Menus

1760 Series Component Waveform/Vector Monitor User Manual

3-24

CALIBRATE

These choices are offered when

CALIBRATE

is selected:

H

CAL SIG

Turns the 1 V internal calibrator signal

on/off.

H

GAIN CAL

Turns on the gain calibration adjust-

ment on/off.

Additional choices appear when

CAL SIG

or

GAIN CAL

is

turned on:

CAL SIG ON

H In waveform mode, the user can set the calibrator

frequency to

100 KHZ

or to

FSC OSC

(

FSC OSC =

3.58 MHz for NTSC/4.43 MHz for PAL).

Dual-stan-

dard monitors also have the option of choosing

NTSC FSC

(3.58 MHz)

or

PAL FSC

(4.43 MHz).

H In vector mode, the user can set the calibrator

frequency, for dual-standard instruments only, to

NTSC FSC

(3.58 MHz)

or

PAL FSC

(4.43 MHz).

H In both waveform and vector modes, selecting

NTSC

FSC

or

PAL FSC

temporarily switches the instrument

to the appropriate video standard. The standard is
restored when changing the selection or when
exiting the

CALIBRATE

menu.

GAIN CAL ON

H In waveform mode, the left bezel control adjusts

instrument vertical gain

(V CAL)

and the center bezel

control adjusts instrument horizontal gain

(

H CAL).

Once gain has been adjusted, two reset options
appear on screen. If either of these resets is se-
lected, a box flashes momentarily around the on-
screen label to acknowledge that reset has occurred.

RESET VCAL

cancels any vertical gain adjustments