Galaxy GHDX2-2430S/R-24F4D Installation and Hardware Reference Manual

5-4

Subsystem

Maintenance

BBU Module

The BBU can be installed after the initial

installation procedure or replaced if a previously

installed BBU module is faulty or fails to hold its

charge.

Main Board

If the controller module in a single controller

model fails, it is necessary to power the system

down and replace the controller.

5.2.2

Notes on Controller Module Maintenance

ƒ

The controller module contains a DIMM module and a BBU module. When

replacing the controller module, these components can be removed and used

on the new controller module if they are functioning normally.

ƒ

When replacing the controller module, you must remember that the

controller board is one of the most sensitive components in the subsystem.

All previously stipulated safety precautions (see Chapter 2) must be strictly

adhered to. Failure to adhere to these precautions can result in permanent

damage to the controller board, resulting in timely delays.

5.2.3

Removing the Controller Module

CAUTION!

1.

It is recommended to apply Galaxy certified memory modules to avoid

compatibility issues.

2.

The installed BBU and controller module must be removed prior to installing

new memory modules. Do this with care. Sensitive components can be

damaged during the process.

3.

The BBU is hot swappable and can be independently swapped from the

controller. However, as a safety precaution and in case your memory still

holds cached data, it is recommended to remove the BBU before handling

the DDR RAM module. If the BBU is supplying power to the memory when

the DDR module is being removed, damage may occur!

To remove the controller module:

Step 1.

Prepare a clean, static-free work pad on which to place the
controller that will be removed from the chassis.