BTTracer Protocol Analyzer User’s Manual

CATC

SW Version 2.20

A menu similar to the one below will open. Your menu may say
"Disable" instead of "Enable."

Step 2

Select "Enable External Trigger Output" (or "Disable
External Trigger Output
" if that is the choice presented.)

If you have chosen "Enable External Trigger
Output
", a small arrow will appear on the right side
of the button. This arrow indicates that a condition
has been set for creating an external output signal.
Choosing "Disable External Trigger" will cause the arrow to disappear.

Enabling High Pulse, Low Pulse or Pulse Toggle Signal Outputs

Once External Trigger Output has been enabled, you can configure the
output signal to one of three formats:

Pulse High - This is the default format. The Pulse High setting causes the
Analyzer to transmit a 5 volt, 16.66 nanosecond signal.

Pulse Low - This format causes the Analyzer to transmit a -5 volt, 16.66
nanosecond signal.

Toggle - This format causes the Analyzer to transmit a signal that will
toggle with each trigger event between a continuous 5 volt signal and a
continuous -5 volt signal.

To configure the output signal,

Step 1

Click the blue dot on an Event button that has a small arrow
attached to it like the one shown above.

A Blue Dot Menu will open. "External Trigger Form" should be a
choice available. If it is not, you will need to choose "Enable External
Trigger
" and then reopen the menu.