Stepped, Burst – Teledyne LeCroy ArbStudio User Manual
Page 45
Operator’s Manual
922244-00 Rev A
37
B
After the generation of the last waveform, ArbStudio starts again
with the first waveform, and then continues.
The sequence can be interrupted at any time by a Force Stop
command.
Stepped
A Force Trigger command activates the generation of the first waveform
present in the generation sequence.
This waveform is repeated for a number of times indicated in the repetition
number.
The instrument output maintains the value imposed by the last
sample of the generated waveform.
A new Force Trigger command activates the generation of
subsequent waveforms.
After the generation of the last waveform, a new Force Trigger
command activates again the generation of the first waveform.
The generation can be interrupted at any time by a Force Stop
command.
Burst
In this mode the first Force Trigger command activates the first waveform
present in the Generation sequence.