Teledyne LeCroy ArbStudio User Manual
Page 80
ArbStudio
72
922244-00 Rev A
The same marker operations are available by right clicking on one of the
markers in the list and selecting the desired function.
DDS
(C
ARRIER
E
DITOR
)
In DDS mode the only available marker identifies a single amplitude value
on the carrier waveform. During the waveform generation, when the
instrument output becomes higher than the marker value, a
leading/trailing edge on the Trigger Out signal is generated.
The Trigger Out signal remains high/low while the output signal remains
higher than the marker amplitude, at which point it switches.
The leading/trailing edge for the Trigger Out signal can be selected on the
Channel Settings Trigger Out tab.
If the selected instrument is configured as the Master, a marker can also
send a synchronous signal on an Expansion Bus line.