Cable Testing in Expert Mode

The Expert Autotest Setup

recommended. Press Next to view the optional measurements you
may want to add to your test. Add them by pressing the name or box.

3

After you have made your optional choices, press Next to allow you
to enable test result analysis functions. It is recommended that you
should choose Manufacturer Specific Compensation when you are
performing Category 6 channel testing. Press Next when done.

4

The next Measurements Setup screen allows you to choose between
Summary ELFEXT data and Complete ELFEXT data storage modes.
If you have storage space and/or you will wish to recertify in the
future using the stored data, the Complete ELFEXT Mode is
recommended. When complete, press Next.

5

The next Measurement Setup Screen allows the user to apply the 3
dB attenuation rule, which provides for special consideration of
measurements where the attenuation is less than 3 dB or in certain
ISO cases, 4 dB. This rule applies to both NEXT and Return Loss in
the ISO case, but only to Return Loss in the TIA case. The
FrameScope 350 automatically sets this choice based upon the
standard chosen previously. It also implements the 4 dB