Agilent Technologies 85225F User Manual

Page 100

Advertising
background image

100

Installation and User’s Guide

3

Verifying System Functionality

This completes the functional verification procedure.

15

For systems with the
Agilent 4284A precision
LCR meter,
open the
junction capacitance
model (juncap.mdl) in
IC-CAP.

a From the IC-CAP/Main window menu bar,

choose File > Examples... .

b In the Directories list of the File Open dialog

box, double-click on the directory
.../examples/model_files.

c In the Directories list of the File Open dialog

box, double-click on the directory
.../model_files/diode.

d In the Files list of the File Open dialog,

double-click on juncap.mdl.

This opens the File Open dialog box.

This opens a list of model files.

(Scroll the list, if needed.) This displays
a list of modeling files in the Files list of
the File Open dialog.

This opens the juncap model window.

The juncap model window contains tab
folders used to interact with the model
data.

16 Select the DUT/Setup.

a Select the DUTs-Setups tab.
b Click the area > cv setup.

This opens the capacitance vs. voltage
setup.

17 Set the LCR meter

instrument options.

a Select the Instrument Options tab.
b Highlight the Cable Length variable and

enter

2

.

This sets the cable length in the LCR
meter instrument options.

18 Calibrate the LCR meter.

a On the 16048D test leads, connect one tee

between the H

pot

and H

cur

connectors.

b Connect the other tee between the L

pot

and

L

cur

connectors. Select the

Measure/Simulate tab.

c Select Calibrate.
d Follow the IC-CAP prompts.

This performs a calibration on the LCR
meter.

The BNC tee part number is 1250-2405.

19 Measure the open circuit

capacitance.

a Select Measure.

This measures the open circuit
capacitance.

20 Observe the results

a After the calibration is complete, observe

the plot.

b Rescale the measured result (displayed in

red) cap.m.

The measured result should be less
than ±10E-15 farads.

* To familiarize yourself with the IC-CAP software, refer to the first three chapters of the Agilent IC-CAP 2004 User’s Guide, model number

85190D.

† To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example

config1.hwd

(the file suffix must be

.hwd

).

‡ To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example

config1.hwd

(the file suffix must be

.hwd

).

**S12 is identical to S21 because the attenuator is assumed to be symmetrical.

To perform the system functional verification test (continued)

Step

Action

Notes