Required tools – Agilent Technologies 85225F User Manual

Page 94

Advertising
background image

94

Installation and User’s Guide

3

Verifying System Functionality

Understanding the System Functional Verification Test

The system functional verification test is a standard IC- CAP measurement
and simulation procedure. The system performs DC and S- parameter
extraction, optimization, and simulation using a fixed 10 dB attenuator as
the device under test (DUT).

The procedure provided in

“Performing the System Functional Verification

Test

” confirms:

the IC- CAP software can communicate with and control the system

instrumentation via GPIB (through the LAN/GPIB gateway, if so
configured)

the system can make measurements and display the results

the IC- CAP software can simulate data

the IC- CAP software can converge the simulated data with the

extracted (measured) data

The system applies forward and reverse current to the attenuator and
monitors voltage at the attenuator’s input and output. IC- CAP then uses
the measured data to model the device- intrinsic resistances and
transmission line delay.

Required Tools

Agilent 85225F performance modeling system

A system controller

*

Agilent 85190A IC- CAP software

Test port cables

Agilent 8490D 10 dB fixed RF attenuator

BNC tee (2)

Agilent 11900A, 2.4 mm male- to- male adapter

Agilent 11900B, 2.4 mm female- to- female adapter

Agilent 11900C, 2.4 mm female- to- male adapter

Agilent 85056A 2.4 mm precision calibration kit, or

Agilent 85056D 2.4 mm economy calibration kit

* For requirements, see

“The System Controller"

on page 59.

† These components are supplied as part of the system.

‡ For CV subsystem verification only

Advertising