B

DC Subsystem Functional Verification Test

For Agilent E5260A/70B DC subsystems:

If you encounter a problem

Refer to the Agilent E5260A/70B precision parametric measurement
solution user’s guide for an explanation of the error codes. See the
section titled “Error Codes” in the chapter titled, “If You Have a
Problem.”

Refer to the troubleshooting information in the Agilent E5260A/70B
precision parametric measurement solution service manual.

To perform the Agilent E5260A/70B precision parametric measurement solution self-test

Step

Notes

1 Disconnect all cables from the

measurement terminals on the
E5260A/70B front panel.

2 Connect the E5260A/70B to line

power.

3 Switch on the instrument line

power switch.

4 Wait 1 hour before continuing to

step 5

.

5 Press Shift > Menu.

6 Move the cursor to SELFTEST and

press Enter. Move the cursor to
EXECUTE and press Enter.

7 Use the arrow keys to select ALL

and press Enter. At the completion
of the test, press Exit three times to
exit the setup menu.

8 To display the results, move the

cursor to SELFTEST and press
Enter. Move the cursor to RESULT
and press Enter.

9 Use the arrow keys to select

FRAME or Slot n.

The test result is displayed for each
item.