Enhancing Measurement Accuracy

To enhance measurement accuracy

Follow these instructions to enhance the accuracy of your
measurements.

To enhance measurement accuracy

Step

Action

Notes

1 Inspect device

connections for foreign
materials or connector
damage.

a Prior to connecting the DUT, inspect the

bias network and semi-rigid RF cable
connectors for foreign materials or
damage.

b Replace any damaged connectors.

Before using connectors for a critical
measurement, inspect the connector
interfaces with a microscope (~20X).
Look for dirt, contaminants, dented or
scratched outer conductor mating
surfaces, and damaged center
conductors.

2 Clean the connectors.

a Remove the systems power cord from the

power supply or place the supply breaker in
the tripped position.

b Move the connectors to a well ventilated

area.

c Use a swab dipped in clean isopropyl

alcohol to cleanse the outer conductor
mating surfaces and the ends of the center
conductors.

d Blow off the remaining alcohol with clean

compressed air.

e Allow adequate time for the alcohol fumes

to disperse before activating the system.

Be careful not to let the alcohol get on
the insulator bead, as this may damage
the bead.

Be careful not to exert too much force
on the center conductors, as they may
be damaged.

Compressed air can reduce the
temperature of connectors
dramatically, and this can have a
significant effect upon the performance
of calibration and verification
components.

If the connector components being
cleaned are to be used in a critical
measurement application, allow the
temperature of these components to
stabilize prior to use.

3 Ensure proper system

warm-up time.

a Switch on line power to all of the system

components.

b Allow at least 1 hour warm-up prior to

making measurements.

Follow the procedure in

“To switch on

power to the system"

on page 87.

Do not switch off line power to the
system unless the system will not be
used for an extended period of time.

4 Ensure a proper and

constant temperature in
the measurement
environment.

a Operate the system within an ambient

temperature range of 25

°

C, ±5

°

C.

b After system measurement calibration,

hold the ambient temperature of the
measurement environment to ±1

°

C of the

ambient temperature at the time of
calibration.

Install heating and cooling systems as
necessary to maintain proper ambient
temperature in the measurement
environment.