Agilent Technologies 85225F User Manual

Page 67

Advertising
background image

Installing the System

2

Installation and User’s Guide

67

3 Unload the system

a Remove the two brace bolts attaching the

side brace assembly to the bottom pallet.

b Remove the side brace assembly.
c Lift the hinged slat and remove the ramp

anchor bolt.

d Place one end of the ramp on the pallet

ramp ledge.

e Insert the ramp anchor bolt and fold down

the hinged slat.

f

Fold down the ramp’s end flap.

To secure the ramp, you may place long
wood screws through the ramp and
into the ramp ledge.

g Ensure that the rack cabinet leveling feet

are retracted and that the cabinet casters
are rolling freely.

h Roll the system down the ramp using

extreme care.

i

Carefully roll the rack toward its prepared
place within the measurement
environment.

In case the system must be moved in
the future, retain and reuse these
packing materials. You can also
purchase replacement packing
materials from Agilent Technologies.

To unpack the shipment crate containing the rack cabinet (continued)

Step

Action

Notes

WA R N I N G

A racked system is tall and top-heavy. It is easy to tip the rack
over while moving it, which could result in injury or death.
Unloading the system safely requires the participation of four
persons exercising care so as not to topple the rack cabinet. Do
not stand in front of the rack as it rolls down the ramp.