To prepare the installation site, Table 16 environmental requirements, Table 17 electrical requirements – Agilent Technologies 85225F User Manual

Page 64: Table 16. environmental requirements, Table 17. electrical requirements, Environmental requirements, Electrical requirements

Advertising
background image

64

Installation and User’s Guide

2

Installing the System

To prepare the installation site

Follow these steps to prepare the site for system installation.

C A U T I O N

This product is designed for indoor use in Installation Category II
and Pollution Degree 2 per IEC 61010-1 and 664 respectively.

To prepare the installation site

Step

Notes

1 Ensure that your installation site meets the

environmental requirements.

Environmental requirements (temperature,
relative humidity, altitude, and clearance) are
listed in

Table 16

.

2 Ensure that your installation site meets the

electrical requirements.

Electrical requirements are listed in

Table 17

.

Table 16 Environmental Requirements

Environmental Parameter

System Requirement

Temperature

+0

°

C to +45

°

C (+32

°

F to +113

°

F)

* †

* Install air conditioning and heating as needed to achieve the required ambient temperature range

.

† Accuracy-enhanced measurement performance is specified at an ambient temperature range of

+25

°

C ±5

°

C. After calibration, hold the ambient temperature of the measurement environment to

±1

°

C of the ambient temperature at the time of calibration.

Relative humidity

Maximum 80% for temperatures up to 31

°

C

decreasing linearly to 50% at 40

°

C

Altitude

Up to 3000 meters (approximately 10000 feet)

Clearance (behind and above rack)

‡ Required to ensure the extractor fans can properly ventilate the system.

15 centimeters (6 inches) minimum

Table 17 Electrical Requirements

Electrical Parameter

System Requirement

Supply capability

100/120 volts, 2000 VA
200/240 volts, 2000 VA

Circuit sharing

Do not connect air conditioning or
motor-operated equipment to the same ac
circuit supplying line voltage to the system.