Network analyzer system performance, Maximum output power, Dynamic range – Agilent Technologies 85225F User Manual

Page 142: Maximum output power dynamic range

Advertising
background image

142

Installation and User’s Guide

G

Network Analyzer Performance Specification Summary

Network Analyzer System Performance

The following specifications describe the system performance of the
Agilent E8364B PNA Series vector network analyzer Option 014 and UNL
configuration for measurements between 45 MHz and 50 GHz.

Calibration Kit

Agilent 85056A, 2.4 mm precision, with sliding loads

Cables

Agilent 85133F, 2.4 mm flexible test port cable set

Calibration Type

Full 2- port with sliding loads

Temperature

Range

25

°

C ±5

°

C

Warm-up Time

0.5 hour

Maximum Output Power

Dynamic Range

System dynamic range is calculated as the difference between the noise
floor and the source maximum output power. Reflection measurements are
limited by directivity. Therefore, system dynamic range only applies to
transmission measurements.

Frequency Range (GHz)

0.045 to 10.0

10.0 to 20.0

20.0 to 30.0

30.0 to 40.0

40.0 to 45.0

45.0 to 50.0

Maximum output power

+3 dBm

0 dBm

4 dBm

8 dBm

11 dBm

17 dBm

Frequency Range (GHz)

0.045 to 0.5

0.5 to 2.0

2.0 to 10.0

10.0 to 20.0 20.0 to 30.0 30.0 to 40.0

40.0 to 45.0 45.0 to 50.0

System
dynamic
range

92 dB

117 dB

120 dB

119 dB

109 dB

105 dB

102 dB

95 dB