Installation and User’s Guide

A

Enhancing Measurement Accuracy

If you encounter a problem

Inspect the connectors on the load, open, short, and through standards
and the connectors on the bias networks and the semi- rigid cables.

Clean the connectors if necessary. See step 2 in the section titled

“To

enhance measurement accuracy"

on page 112.

If connectors are damaged, replace the standard or cable.

Ensure that the standards meet their published specifications.

5 Calibrate the network

analyzer through the bias
networks, cables, and
adapters using the
standard open, short,
load, and through.

a Follow the prompts to connect the standard

open to the Port 1 bias network RF/DC
OUTPUT.

b Click Measure.
c Follow the prompts to connect the standard

short to the Port 1 bias network RF/DC
OUTPUT.

d Click Measure.
e Follow the prompts to connect the standard

broadband load to the Port 1 bias network
RF/DC OUTPUT.

f

Click Measure.

g Repeat actions a through f for the Port 2

calibration measurements.

h Follow the prompts to connect the standard

through between the Port 1 and Port 2 bias
networks RF/DC OUTPUT. connectors.

i

Click Measure.

j

When the measurements are done, in the
Standards Measured dialog box, click Next.

k In the Guided Calibration Completed dialog

box, choose the No, Finish Now radio
button and click Finish.

6 Save the calibration and

instrument state data to
the C:/Program Files/
Agilent /Network
Analyzer/Documents

folder.

a In the File menu, select Save As... .
b In the Save As dialog box, using the

keyboard or by clicking Edit File Name,
enter a file name (for instance,

MY_CAL.CST

), then click OK.

This saves the calibration and
instrument state data as a file (named
for instance, MY_CAL.CST) in the
network analyzer’s operating system
C:/Program Files/Agilent/
Network Analyzer/Documents
folder.