To remove or replace a system component – Agilent Technologies 85225F User Manual

Page 105

Advertising
background image

Servicing the System

4

Installation and User’s Guide

105

To remove or replace a system component

Follow these steps to remove a system component for periodic component
calibration, service, or repair.

Follow the steps in reverse order to replace a system component.

WA R N I N G

These servicing instructions are provided for use by qualified
personnel only. To avoid electrical shock, do not perform any
servicing unless you are qualified to do so.

WA R N I N G

The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the product from all voltage
sources before opening covers or removing parts.

To remove a system component

Step

Note

1 Turn off all components.

2 Disconnect the system from line

power.

3 Disconnect all cables from the

instrument to be removed.

4 Prepare a surface near the rack

cabinet to place the removed
system component.

5 Lower the stabilizing legs on the

rack cabinet.

6 From the front of the rack cabinet,

remove the four screws (2 screws
on each rack mount) attaching the
instrument’s rack mount and
handle kit to the rack cabinet.

7 As another person steadies the

rack cabinet, hold the instrument
by the rack mount handles and
slowly pull forward.

Some components may weigh more
than 50 pounds and may require
more than one person to remove
safely.

8 Save the rack mount screws by

reinserting them in the rack cabinet
frame nuts.