Step 13 – Agilent Technologies 85225F User Manual

Page 88

Advertising
background image

88

Installation and User’s Guide

2

Installing the System

13 On the 4156C, press System >

[MISCELLANEOUS], highlight the 4156C
value in the GPIB ADDRESS field, and press
19 > Enter.

This sets the correct system GPIB address
(19) for the 4156C.

14 On the E5260A/70B, move the cursor to

CONFIG, press Enter, move the cursor to
ADDRESS, press Enter. Use the arrow keys
to set the address to 19 and press Enter.

This sets the correct system GPIB address
(19) for the E5260A/70B.

15 Switch on the Agilent E8364B PNA Series

vector network analyzer line power.

16 On the Agilent E8364B PNA Series vector

network analyzer, from the Main dialog,
select System > Configure > SICL/GPIB. In
the SICL/GPIB dialog GPIB group box, select
the Talker/Listener radio button and select
16 in the Address scroll list.

This sets the correct system GPIB address
(16) for the E8364B.

17 If present, switch on the LAN/GPIB gateway

line power.

Refer to the LAN/GPIB gateway documentation
for instruction on installation and configuration.

18 Switch on the computer line power.

19 Allow the system to warm up for one hour.

* If the system is to be used with an autotransformer, ensure that the common terminal is connected to the

neutral (grounded) side of the power source.