Network Analyzer Performance Specification Summary

G

Installation and User’s Guide

143

Measurement Port Characteristics

Measurement Uncertainty

Measurement uncertainty curves utilize an RSS (Root Sum Square) model
for the contribution of random errors such as noise and typical connector
and test set switch repeatabilities. These are combined with a worst- case
model for the contributions of dynamic accuracy and residual systemic
errors.

Curves show the worst- case magnitude and phase uncertainty for
reflection and transmission measurements, using the specified cal kit, with
10 Hz IF bandwidth, and no averaging during the measurement.

Reflection Measurements

Frequency Range (GHz)

Residual

0.045 to 2.0

2.0 to 20.0

20.0 to 40.0

40.0 to 50.0

Directivity

42 dB

42 dB

38 dB

36 dB

Source match

41 dB

38 dB

33 dB

31 dB

Load match

42 dB

42 dB

37 dB

35 dB

Reflection tracking

±(0.001 + 0.2 dB/

°

C)

±(0.008 + 0.2 dB/

°

C)

±(0.020 + 0.3 dB/

°

C)

±(0.027 + 0.4 dB/

°

C)

Transmission tracking

±(0.019 + 0.2 dB/

°

C)

±(0.053 + 0.2 dB/

°

C)

±(0.114 + 0.3 dB/

°

C)

±(0.215 + 0.4 dB/

°

C)