Agilent Technologies 85225F User Manual

Page 6

Advertising
background image

6

Installation and User’s Guide

safety standards of design, manufacture, and intended use of the product.
Agilent Technologies, Inc. assumes no liability for the customer’s failure to
comply with these requirements.

For additional safety precautions, including precautions for making device
measurements in a floating ground configuration, see

“To ensure your

safety while using the system"

on page 76.

WA R N I N G

This is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the mains supply cord). The mains plug shall
be inserted only in a socket outlet provided with a protective earth
contact. Any interruption of the protective conductor inside or
outside of the product is likely to make the product dangerous.
Intentional interruption is prohibited.

WA R N I N G

If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used only in a
normal condition (in which all means for protection are intact) only.

WA R N I N G

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE. Do not operate
the instrument in the presence of flammable gases or flames.

WA R N I N G

DO NOT REMOVE THE INSTRUMENT COVER. Operating personnel
must not remove instrument covers. Component replacement and
internal adjustments must be made only by qualified service
personnel. Instruments that appear damaged or defective should be
made inoperative and secured against unintended operation until
they can be repaired by qualified service personnel.

WA R N I N G

Installing additional instruments may destabilize the rack cabinet.

WA R N I N G

Installing additional instruments into the cabinet electrical system
could produce excessive leakage current. If the protective earth
conductor is interrupted or faulted, the user risks serious injury or
death.

WA R N I N G

Prior to adding any additional instruments, review all wiring and
cooling capabilities to verify adequate design margins for normal and
under single fault conditions.

Advertising