Agilent Technologies 85225F User Manual

Page 82

Advertising
background image

82

Installation and User’s Guide

2

Installing the System

Agilent 4156C Systems with Agilent 41501B Expander Box

To connect the bias networks

Step

Action

Notes

1 For Agilent 4156C with

41501B expander box
systems, refer to the
following figure.

2 Connect the triaxial

cables from the 4156C to
the bias networks.

a Connect the triaxial cables from the 4156C

HRSMU1 FORCE and SENSE connectors to
the DC FORCE and DC SENSE connectors
on the port 1 bias network.

b Connect the triaxial cables from the 41501B

HPSMU FORCE and SENSE connectors to
the DC FORCE and DC SENSE connectors
on the port 2 bias network.

From the rear of the system, route the
cable through the feedthrough panel.

The port 1 bias network is the 11612V
K11.

The port 2 bias network is the 11612V
K21.

Refer to

Figure 20

, “Rear Panel Wiring

Diagram,” on page 53.

3 Connect and route the

triaxial cable from 4156C
HRSMU3 FORCE
connector to the port 2
bias network.

a From the rear of the system, connect one

end of the triaxial cable to the GNDU
connector on the 41501B rear panel.

b From the rear of the system, route the cable

through the upper feedthrough panel.

c Connect the triaxial cable to the port 2 bias

network GNDU connector.

The triaxial cable model number is
16494A Option 002.

Leave the GNDU connector on the port
1 bias network open.

Refer to

Figure 3

, “Front Panel

Connections with Agilent 4156C,” on
page 25.