Periodic system component calibration – Agilent Technologies 85225F User Manual

Page 117

Advertising
background image

Enhancing Measurement Accuracy

A

Installation and User’s Guide

117

Periodic System Component Calibration

A complete calibration verifies that the system components meet their
individual performance specifications.

The calibration interval depends on the level of system use. Agilent
Technologies recommends an initial cycle of 6 to 12 months. Thereafter,
adjust the cycle based on the recalibration results.

A complete system calibration consists of the following tests:

Agilent E8364B PNA Series vector network analyzer performance tests

*

Agilent 4156C precision semiconductor parameter analyzer performance

verification

Agilent E5260A/70B precision parametric measurement solution

performance verification

Agilent 4284A precision LCR meter performance tests

Agilent 35670A dynamic signal analyzer performance verification

Agilent E5250A low leakage switch mainframe performance tests

See the individual instrument documentation for instructions on
performing the required instrument verification tests.

* Refer to the Agilent E8364B PNA Series vector network analyzer service guide for required performance tests.

The service guide (part number E8364-90026) is available at www.agilent.com in PDF format.

† Refer to the chapter titled “Performance Verification” in the Agilent 4156C precision semiconductor

parameter analyzer service guide, included with the Agilent 4156C documentation.