Performing a, Coaxial system measurement calibration – Agilent Technologies 85225F User Manual

Page 115

Advertising
background image

Enhancing Measurement Accuracy

A

Installation and User’s Guide

115

Performing a Coaxial System Measurement Calibration

Complete the following steps to perform a coaxial system measurement
calibration and increase the accuracy of your device measurements. Modify
these instructions as necessary to perform an in- fixture or on wafer
measurement calibration.

To perform the system measurement calibration

Step

Action

Note

1 Preset the network

analyzer.

On the network analyzer, press Preset.

2 Open the Calibration

Wizard and choose a
guided calibration.

a In the network analyzer’s Calibration menu,

select Calibration Wizard... .

b In the Calibration Wizard dialog box, select

Use Mechanical Stds... in the Guided
Calibrations group box.

3 Choose the DUT

connector types.

a In the Guided Calibration: Select DUT

Connectors dialog box, select the
appropriate DUT connector type for the
connections the Port 1 and Port 2 bias
networks.

b Click Next.

4 Choose the calibration kit. a In the Guided Calibration: Select Version 2

Cal Kits dialog box, select the appropriate
calibration kit (for example 85056D) for the
Port 1 and Port 2 bias network.

b Click Next.