D

CV Subsystem Functional Verification Test

Understanding the CV Subsystem Functional Verification Test

Use the procedure provided in

“Performing the CV Subsystem Functional

Verification Test

to manually confirm the functionality of the

Agilent 4284A precision LCR meter. This procedure does not verify
performance to specification. This procedure confirms that the precision
LCR meter is ready for performance verification and/or operation by
confirming the following hardware functionality:

memory card read/write test

LED display test

LCD display test

handler I/F test

scanner I/F EEPROM read/write test

scanner I/F I/O test

bias current I/F I/O test