Performing a calibration check, Recalibration procedure – West Control Solutions Pro-EC44 User Manual

Page 80

Advertising
background image

Pro-EC44 2-Loop Graphical Profile Controller & Recorder

Pro-EC44 Product Manual - 59540-1 October 2013

Page 73

Performing a Calibration Check

1. Setup input 1 for the input signal type to be checked.

2. Power up the instrument and correctly connect the signal source.

Leave powered up for at least five minutes for RTD and DC linear inputs, and at least
30 minutes for thermocouple inputs.

3. After the appropriate delay for stabilisation, check the calibration at a number of

cardinal points by applying the appropriate input signal.
The observed readings should be within the tolerances stated in the specifications
(see page 245).

4. Test the other signal types as above if required.

5. Repeat the process for input 2 if fitted.

Recalibration Procedure

For each process input, recalibration is carried out in five phases as shown in the table
below; each phase corresponds to a basic input type.

Note: The 50mV calibration phase MUST always be calibrated before calibration
of the thermocouple input.

INPUT CALIBRATION PHASES

Type

Signal

(<0.05% error)

Cable Type

Input 1 Terminals Input 2 Terminals

+

-

+

-

DC milli-volt

50 mV

Copper Wire

2

3

6

7

DC voltage

10 V

Copper Wire

2

3

6

7

DC milliamps

20 mA

Copper Wire

3

1

7

5

RTD

200 ohm

Copper 3-Wires

1

2 & 3

5

6 & 7

Thermocouple

0ºC K type source

K Thermocouple Wire

2

3

6

7

1. For optimum accuracy, leave the instrument power-up for >30 minutes to warm up before

beginning the calibration, and then toggle the power off/on to restart the instrument.

2. During the power-

up “splash screen”, press

D

and

R

together until the Input 1

Calibration Status screen is displayed.

3. Correctly connect the 1

st

phase signal (50mV), then press

R

to select the first phase

4. Press

D

+

U

to initiate the calibration.

5.

During calibration the message “50mV DC Input Calibrating” will display for a few
seconds. This should be followed by the “Calibration Successful” confirmation.