4 water reservoir, E-m-hg2-s-v2.1 – ROTRONIC Hygrogen2 User Manual

Page 11

Advertising
background image

E-M-HG2-S-V2.1

Rotronic AG

Bassersdorf, Switzerland

Document code

Unit

HygroGen2: Humidity and Temperature Generator

with AutoCal/AutoCal+, Remote Control and

Range Extensions options.

Instruction Manual

Instruction Manual for Software

Version 2.1

Document Type

Page 7 of 75

Document title

2.4 Water Reservoir

The HygroGen2 requires 50-80ml of distilled water in the internal reservoir. Insufficient water will

deactivate the humidification system. As long as extreme care is taken not to overfill the water reservoir,

the HygroGen2 may be filled and emptied either with the unit powered up or shut down.

To fill the HygroGen2

1. Insert the tube of the dosing syringe, with the plunger fully depressed, 10mm into the water

port and extract any residual water by drawing out the plunger.

2. Depress the collar on the water port (Figure 1(4) above) and remove the dosing syringe tube.

Wipe up any spilt water with an absorbent cloth immediately.

3. Fill the dosing syringe with 80ml of distilled water. Insert the end of the dosing syringe tube

firmly back into the water port and discharge the syringe steadily. (If water leaks during this

operation the filler tube is not correctly inserted in the water port.)

4. Depress the collar on the water port to remove the dosing syringe tube. Wipe up any spilt water

with an absorbent cloth immediately.

5. A water level indicator is shown in the Settings Screen (see

section 3.4.5

).

Caution: Do not overfill the HygroGen2

Overfilling of the unit will result in poor humidification and, in extreme cases, water may be pumped

into the chamber. If the HygroGen2 is overfilled, the humidifier should be fully emptied and then refilled

with 50-80ml of water (as described above). If there is water present in the chamber of the HygroGen2,

this should be removed using an absorbent cloth. Ensure the HygroGen2 is switched off and the mains

power unplugged before manually drying the chamber.


Figure 3: How to fill and empty the Water Reservoir

© 2014; Rotronic AG

E-M-HG2-S-V2.1

Advertising