4 range extensions, E-m-hg2-s-v2.1 – ROTRONIC Hygrogen2 User Manual

Page 46

Advertising
background image

E-M-HG2-S-V2.1

Rotronic AG

Bassersdorf, Switzerland

Document code

Unit

HygroGen2: Humidity and Temperature Generator

with AutoCal/AutoCal+, Remote Control and

Range Extensions options.

Instruction Manual

Instruction Manual for Software

Version 2.1

Document Type

Page 42 of 75

Document title

4.4 Range Extensions

Range extension allows the unit to operate down to -5°C and 2%rh and up to 99%rh (between 10°C and

45°C). The unit is able to reach 90%rh above 45°C and below 10°C.

Please be aware that it may take a long time to reach set-points at these extremes of temperature and

humidity, and the unit may need to be individually adjusted to optimise performance. Ambient

conditions will affect the performance of the HygroGen2 at low temperatures. Units that have not been

serviced recently may not achieve these extremes. Control HygroClip2s that are not specifically

calibrated for low temperature & RH operation may require adjustment. Expert adjustment of the

control PID parameters may also be required. PID access is via the Service Menu, which is locked down

for standard operation. For further information please contact

[email protected]

.

With the Humidity Range Extension the HygroGen2 may be further configured to allow set-points

between 0 and 99.9% RH (maximum 90%RH below 10C). This is done from the Service menu on the

Service screen. The performance and stability of the instrument as it approaches these extreme

conditions is not guaranteed. Furthermore, the accuracy of the control probe needs to be borne in mind.

© 2014; Rotronic AG

E-M-HG2-S-V2.1

Advertising