E-m-hg2-s-v2.1 – ROTRONIC Hygrogen2 User Manual
Page 51
E-M-HG2-S-V2.1
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroGen2: Humidity and Temperature Generator
with AutoCal/AutoCal+, Remote Control and
Range Extensions options.
Instruction Manual
Instruction Manual for Software
Version 2.1
Document Type
Page 47 of 75
Document title
Transit
Use “Transit” to shut down and condition the HygroGen2 if the unit is to be transported or stored for
an extended period. The HygroGen2 will ask for the unit to be emptied of water. It will isolate the
humidifier from the chamber and then cool/dry the chamber to ensure excess water is removed. This
process may take several minutes.
Transit Procedure
1. Press the “Power Off” button on-screen.
2. Select “Transit”.
3. Empty water as prompted on-screen.
4. Allow the HygroGen2 to shut down. The HygroGen2 will isolate the humidifier and cool/dry to
ensure excess water is removed. This may take several minutes.
HygroGen2 will not switch off if the water is not emptied as requested!
DO NOT switch off at the wall before the HygroGen2 has shut down. Failure to do this may
cause software corruption.
© 2014; Rotronic AG
E-M-HG2-S-V2.1