3.4.4 Edit Program Screen

1) Graphic Indicator

2) Ramp/Soak control

3) Temperature set-point

4) Humidity set-point

5) Step duration

6) Insert Step

7) Delete Step

8) Scroll to view Steps

9) Program title

10) Temperature tolerance

11) Humidity tolerance

12) Temperature Only

13) Programmer

Select a Program to Edit

Select the desired program by clicking on the program button and entering the desired program

number.

Edit a Program Name

To edit the program name click on it (USB keyboard required) (item 9).

Edit a Step

Each program is made of steps. Each step can be edited. Enter the desired step set-points and duration

(items 3, 4 & 5). A graphical representation of the step is shown to help visualise the sequence.

Insert or Delete a Step (items 6 & 7)

Use the insert and delete buttons to add and remove steps. If there are more than 7 steps in the program

use the scroll bar (item 8) to view.

Ramp or Soak (item 2)

Set temperature and humidity to either ramp or soak using item 2. A soak is where the HygroGen2

maintains the temperature and humidity in the chamber constant for the duration of the step, within

the tolerances set (items 10 & 11). A ramp is where the HygroGen2 changes the selected set point at a

rate determined by the subsequent set-point value and the current step duration.

Figure 11: Edit Program