ROTRONIC Hygrogen2 User Manual

Page 73

Advertising
background image

D2: HygroGen2 Uncertainty Framework - External RH Probe

Input values:

23

°C

80

% RH

Temperature sensitivity

4.96 % RH/deg C

Reference Uncertainty source

Distribution

Value

Units

Sensitivity

Equiv RH Divisor One SD Squares

1

Uncertainty of reference probe

normal

0.900000

%rh

1.000000

0.900000 2.00

0.450

0.203

2

Repeatability of humidity

normal

0.100000

%rh

1.000000

0.100000 1.00

0.100

0.010

3

Reference drift

rectangular

1.000000

%rh

1.000000

1.000000 1.73

0.577

0.333

4

Reference resolution

rectangular

0.050000

%rh

1.000000

0.050000 1.73

0.029

0.001

5

Chamber temperature gradients

rectangular

0.020000

°C

4.960000

0.099200 1.73

0.057

0.003

6

Gradients through probe stem

rectangular

0.100000

°C

4.960000

0.496000 1.73

0.286

0.082

7

Temperature stabilisation

rectangular

0.005000

°C

4.960000

0.024800 1.73

0.014

0.000

8

Repeatability of temperature

normal

0.050000

°C

4.960000

0.248000 1.00

0.248

0.062

9

Humidity stabilisation

rectangular

0.050000

%rh

1.000000

0.050000 1.73

0.029

0.001

10

Humidity fluctuations

rectangular

0.050000

%rh

1.000000

0.050000 1.73

0.029

0.001

11

Instrument Under Test (IUT) Resolution

rectangular

0.050000

%rh

1.000000

0.050000 1.00

0.050

0.003

12

Standard deviation of Ref readings

normal

0.050000

%rh

1.000000

0.050000 1.00

0.050

0.003

13

Standard deviation of IUT readings

normal

0.100000

%rh

1.000000

0.100000 1.00

0.100

0.010

Reference

Sum of squares

0.71

1

From calibration certificate - check uncertainty at input values

Standard uncertainty (√

sum of squares)

0.843

2

From Standard Deviation of 10 repeated calibration runs

Expanded uncertainty (95 % confidence)

1.69

3

Drift between two successive calibrations of the reference instrument

4

Resolution from manufacturer‘s specification

5

Measured temperature gradients from the ROTRONIC HygroGen2 Service

6

Stem conduction

7

Measured stability from the ROTRONIC HygroGen2 Service

8

From Standard Deviation of 10 repeated calibration runs

9

Resolution of ROTRONIC HygroGen2 Control Clip

10

Resolution of ROTRONIC HygroGen2 Control Clip

11

Resolution from manufacturer‘s specification

12

Standard Deviation of Ref readings during calibration

13

Standard Deviation of IUT readings during calibration