HW4 activation

HygroGen2 version 2.0 installs HW4 version 3.2.0. If the code you noted in

section 3.2.1 step i(c)

is

not version 3.x, proceed to step (a) below. Otherwise, proceed to step (h).

a. To use HW4, you have 3 options. Either you can downgrade to HW4 v2.4, buy and upgrade to

HW4 v3 or buy the AutoCal Enhanced Feature set which includes HW4 v3. Please note, a

standard PC HW4 V3 licence will not work.

To downgrade to HW4 V2:

b. On the root of the USB installation stick is a file “HW4 v2.zip”. On a PC, right-click this file and

click “Extract to…" and choose the root directory of a USB stick, e.g. D:\ (i.e. not a subfolder e.g.

D:\HW4 v2). You can use the Installation USB stick itself. A folder “HW4 Update” will be created

in the root directory.

c. Safely remove the USB stick from the PC.

d. On the HygroGen2, go to the Settings screen, then select “Update Software” from the

“Updates” drop-down menu.

e. Select OK to boot the HygroGen2 into configuration mode.

f. Once the HygroGen2 is in configuration mode, insert your USB disk and select “Update HW4”.

NB: If “Update HW4” is greyed out, the folder “HW4 update” has not been found in the root of

the inserted USB stick.

g. Press OK to reboot.

Having obtained an HW4 V3 licence key or downgraded to HW4 V2.4, now enter the HW4

code:

h. On the HygroGen2, press the HW4 button. Select “Start HW4” and enter the code.

3.2.2 Installation of HygroGen2 version 2.1 (from version 2.0)

Version 2.1 is installed via the “Updates” drop-down menu. Go to the Settings screen and select

“Update Software”. Press OK to restart the instrument into Configuration Mode.

a. Download the software update from www.rotronic.com to a PC.

b. Open up the zip file on a PC and copy the folder “Controller Update” to the root of a USB stick.

c. Remove the USB stick from the PC and insert into any USB port on the HygroGen2

d. Once the HygroGen2 detects a USB stick with the controller software, the “Update HG” button

should become enabled. Press this.

e. Follow the on-screen instructions, accepting the default directory locations and accepting the

software terms and conditions.