3.3 Configuration Mode Settings

Certain settings on the HygroGen2 must be made in Configuration Mode. From Standard Operation

Mode, go to the Settings screen, then the “Settings” drop-down menu and select either “Network”,

“Date and Time”, “Touchscreen”, or “Update Software”. All four options effectively do the same thing

(i.e. reboot the HygroGen2 into Configuration Mode); press “OK” to restart the HygroGen2.

3.3.1 Network Settings

Prior to machine serial number VCT-HG2-1281, the network port on the back of the HygroGen2 was

physically disabled. Since this unit, and for every unit that has received a service since January 2013, the

network port has been physically enabled, but its functionality has been disabled in software.

As of v2.1, there is an option to enable the functionality, which is required for the Remote

Control/Support Enhanced Feature. If you do not intend to use the Remote features, the Network

functionality can still be enabled, to no disadvantage, but some organisations may stipulate it is turned

off before the instrument can be used.

By default, when v2.1 is installed, the network port functionality is disabled. To enable it, select

“Network” from the “Settings” drop-down menu from within the Configuration Mode “Settings and

Updates” screen.

Figure 4: Configuration Mode

© 2014; Rotronic AG

E-M-HG2-S-V2.1