HygroGen2: Humidity and Temperature Generator

with AutoCal/AutoCal+, Remote Control and

Range Extensions options.

Instruction Manual

Instruction Manual for Software

Version 2.1

Document Type

Page 5 of 75

Document title

2.3 HygroGen2 Assembly

1)

Power switch

2)

USB ports

3)

Touch screen controller

4)

Water port

5)

Desiccant cell

6)

Chamber door

7)

Probe aperture bung

8)

Handle

1)

Sample loop return

2)

Sample loop outlet

3)

USB ports

4)

DVI monitor interface

5)

Ethernet socket

6)

Type label

7)

Fan outlet

8)

Electrical mains connection

9)

Power switch

HygroGen2 Chamber and Door

The HygroGen2 does not come supplied with a door. There are various standard and customized

versions available, depending on the intended use for the instrument, the use of an external reference

(see

Appendix A1.2

) and the instruments under test. Figure 1(6) above shows the 15mm 5-port version

(part code: HG2-D-11111). Other options are listed in

Appendix B4

.

To access the HygroGen2 chamber, remove the door. Grip firmly and twist the door anti-clockwise. In

the event that this proves difficult, cool the chamber first. Occasional application of silicone grease

around the door “O” ring seal will ease movement.

Figure 1: Front of HygroGen2

Figure 2: Rear of HygroGen2