5 general operational considerations, 1 water in the chamber, 2 below ambient temperature operation – ROTRONIC Hygrogen2 User Manual

Page 47: E-m-hg2-s-v2.1

Advertising
background image

E-M-HG2-S-V2.1

Rotronic AG

Bassersdorf, Switzerland

Document code

Unit

HygroGen2: Humidity and Temperature Generator

with AutoCal/AutoCal+, Remote Control and

Range Extensions options.

Instruction Manual

Instruction Manual for Software

Version 2.1

Document Type

Page 43 of 75

Document title

5 General Operational Considerations

5.1.1 Water in the Chamber

Before using the HygroGen2, check that there is no visible water inside the chamber. If the last condition

run by the HygroGen2 was at high humidity and the unit not correctly shut down (see below),

condensation may have formed. Remove excess water using absorbent tissue and initially run the

HygroGen2 at an elevated temperature (40°C) and low RH set-point to dry the system before

calibrations are run.

5.1.2 Below Ambient Temperature Operation

If the chamber is cooling (or maintaining a temperature below ambient), then the heat pump assembly

inside the chamber will necessarily be below the chamber temperature. If the humidity in the chamber

is high, the cool heat pump assembly can bring the local air temperature below the dew point, causing

condensation to occur on the internal heat pump assembly. If run under these conditions for extended

periods, condensation will collect in the chamber. This will not affect chamber performance within

normal operating ranges (although it will increase the time to achieve low humidity). Water in the

chamber can be removed using an absorbent tissue. Ensure the HygroGen2 is switched off before

manually drying the chamber.

When operating the unit at high humidity and chamber temperatures of more than a few degrees below

ambient, the following procedure will minimise condensation in the chamber during stabilisation:

1. Set the set-point temperature to approximately ambient.

2. Set the set-point humidity to 10%rh and allow the HygroGen2 to stabilise.

3. Set the humidity control to Manual (as described in Automatic/Manual Mode in

section 3.4.1

).

4. Apply the temperature set-point to that desired and allow the HygroGen2 to stabilise.

5. Wait for the Humidity reading to fully stabilise (this may take 15-20 minutes).

6. Set the humidity control loop to Automatic.

It is recommended that the chamber door is not opened when the chamber temperature is below the

ambient temperature as water vapour from the atmospheric air introduced will condense out into the

chamber. The dehumidification of the chamber takes longer at lower chamber temperatures.

© 2014; Rotronic AG

E-M-HG2-S-V2.1

Advertising