HygroGen2: Humidity and Temperature Generator

with AutoCal/AutoCal+, Remote Control and

Range Extensions options.

Instruction Manual

Instruction Manual for Software

Version 2.1

Document Type

Page 19 of 75

Document title

When set-point combinations are programmed that resolve a low dew point condition (i.e. 0°C, 5%rh =

-36°C dew point), make sure that the desiccant is dry enough to achieve the desired value. Check the

desiccant condition on the Settings screen (Figure 12) and consult maintenance records to establish

when the desiccant was last changed. Stabilisation time to low RH conditions will be longer than normal

as the chamber and internal surfaces need to be very dry.

Temperature and Humidity Control (items 5 & 9)

The Temperature and humidity control can be set to automatic or off. Automatic mode results in the

unit attempting to adjust to the temperature (or humidity) in the chamber to match the set-point value.

When switched off the HygroGen2 does not control the parameter.

When off, the control buttons are grey.

When on, but not stable, the control buttons are red.

When on and stable, the control buttons are green.