Figure 20: Edit AutoCal Reference Information

Reference calibration information can be entered here, including the last and next-due calibration dates

and any comments.

When an AutoCal run is started, the user is warned should the reference calibration be out of date and

prompted to update the reference information should the reference probe have been swapped out. The

user is further asked to confirm that the operator name to be printed on the certificate does not require

changing. If a probe that has previously been used and has reference information stored on the

HygroGen2 is restored to the instrument, the information will automatically be reapplied.