Document title

8 Servicing and Maintenance

8.1 Control HygroClip Calibration

To remove the control HygroClip for calibration, turn off the HygroGen2 auto Control and remove the

door. Reach into the chamber and unscrew the HygroClip from its mounting connector. Take care not

to cross the threads on the collar when refitting.

To minimise HygroGen2 downtime, alternative calibrated probes can be fitted in place of the existing

probe.

8.2 Desiccant

The HygroGen2 uses a desiccant when lowering humidity in the chamber. The desiccant condition is

shown within the Settings screen. This shows an approximate indication of the dew point within the

dryer cell.

If the HygroGen2 does not reach low RH set-points promptly, check the desiccant condition and

replace as required.

The measured desiccant value is established when the dryer is active and a flow of air from the

desiccant cell is passing over the desiccant sensor; this value is held when the dryer is active to provide

the user with an indication of the desiccant condition even if the dryer is not used. To update the

value, run the HygroGen2 on the lowest RH set-point (5%rh) for at least 15 minutes.

Changing Desiccant

Remove the desiccant cell from the HygroGen2 by pulling firmly on the cell cap (Figure 1(5)). The cell

is held in place securely to maintain an air-tight seal; a gentle twist will help removal. Changing the

desiccant should be undertaken in a ventilated area.

Observing the diagram below (Figure 36), remove the cap and pour out the used desiccant material

and replace with new or regenerated material. Avoid exposing the replacement desiccant unduly to

the atmosphere.