ROTRONIC Hygrogen2 User Manual

Page 74

Advertising
background image

D3: HygroGen2 Uncertainty Framework - Chilled Mirror Reference

23

°C

50

% RH

Temperature sensitivity

3.1

% RH/deg C

Reference Uncertainty source

Distribution

Value

Units

Sensitivity

Equiv RH Divisor One SD Squares

1

Uncertainty of chilled mirror reference

normal

0.200000

°C

3.100000

0.620000

2.00

0.310

0.096

2

Repeatability of humidity

normal

0.100000

%rh

1.000000

0.100000

1.00

0.100

0.010

3

Chamber temperature gradients

rectangular

0.020000

°C

3.100000

0.062000

1.73

0.036

0.001

4

Temperature stabilisation

rectangular

0.005000

°C

3.100000

0.015500

1.73

0.009

0.000

5

Repeatability of temperature

normal

0.050000

°C

3.100000

0.155000

1.00

0.155

0.024

6

Humidity stabilisation

rectangular

0.050000

%rh

1.000000

0.050000

1.73

0.029

0.001

7

Humidity fluctuations

rectangular

0.050000

%rh

1.000000

0.050000

1.73

0.029

0.001

8

Instrument under test (IUT) resolution

rectangular

0.050000

%rh

1.000000

0.050000

1.00

0.050

0.003

9

Standard deviation of Ref readings

normal

0.050000

%rh

1.000000

0.050000

1.00

0.050

0.003

10

Standard deviation of IUT readings

normal

0.100000

%rh

1.000000

0.100000

1.00

0.100

0.010

Reference

Sum of squares

0.15

1

See sheet Dew Point UA

Standard uncertainty (√

sum of squares)

0.385

2

From Standard Deviation of 10 repeated calibration runs

Expanded uncertainty (95% confidence)

0.77

3

Measured temperature gradients from the ROTRONIC HygroGen2 Service

4

Measured stability from the ROTRONIC HygroGen2 Service

5

From Standard Deviation of 10 repeated calibration runs

6

Resolution of ROTRONIC HygroGen2 Control Clip

7

Resolution of ROTRONIC HygroGen2 Control Clip

8

Resolution from manufacturer’s specification

9

Standard Deviation of Ref readings during calibration

10

Standard Deviation of IUT readings during calibration