3.4.3 Programmer Screen

The HygroGen2 programmer function enables the user to automate a sequence of RH and/or

temperature set-points. If the AutoCal function is enabled the Programmer function is only available via

the Home menu (Figure 8 item 1).

The HygroGen2 Programmer function can accommodate 20 programs with up to 200 steps each.

Programs are edited via the Edit Program (Figure 10).

1) Program Name & Number

2) Program Selection

3) Step Selection

4) Run/Stop Program

5) Edit Program

Program Selection (items 1 & 2)

Item 1 identifies the program number and title. You can use the “<<” and “>>” buttons to scroll through

the saved programs.

Step Selection (item 3)

Use the Step Selection buttons (item 3) to select the program step at which the program will start. The

Temperature and Humidity set-points (SP) for the selected step are displayed.

Run/Stop (item 4)

Start the selected program running at the selected step using the “>” button. The total time the program

has been running, and the time left on the current step are displayed. Use the Stop button to halt the

program before it has completed.

Edit Program (item 5)

Use this button to go to the Edit Program screen (see Figure 11).