4.1.6 AutoCal Operation and Best Practices

Once the program and adjustment points are determined, programs are started by pressing the “Play”

(>) button. After a successful run, provided AutoCal has successfully communicated with the probe

under test for every selected sample point at every set-point, AutoCal saves a time-stamped PDF

calibration certificate to a USB drive plugged into the HygroGen2. The same information is stored in a

corresponding .csv file. If, for any reason, communication with the probe under test is interrupted (by

unplugging the probe, for example), only the .csv file will be saved, with the associated data actually

recorded. Included in each file is the program name, set- and adjustment-points, probe serial numbers

and relevant calibration reference information. It is, further, possible to upload a custom logo and other

user and reference information (see

section 4.1.7

). A separate PDF is generated for each probe. NB:

AutoCal will not run unless a USB drive is detected by the HygroGen2.

Before selecting a probe for adjustment, it is recommended to perform an initial assessment calibration

run without adjustment selected. At the end of the run, examine the PDF calibration certificate saved

to the USB drive and determine whether you want to re-run the program with adjustment selected.

At the first adjustment-point to be applied, AutoCal will initially delete any previously saved calibration

points stored on the probe. This is recorded on the certificate. It will then apply the values of the

reference probe as a corrected value. This value is determined at the instant the correction is saved and

is not an average of the previously recorded points as shown in the calibration certificate; these are

shown in the certificate to give the user an estimate of stability.

Please note, for probes with adjustment selected, AutoCal will always attempt to make an adjustment.

If the correction is smaller than the resolution of the probe itself, the adjustment may not be applied.

If, at the end of the run, the probe has acknowledged all instructions it has been sent, the “Completed“

button will turn green on the AutoCal screen and the PDF is saved to the USB drive. Details of the probe’s

progress throughout the run can be viewed via the “Details” button.

Please note, the performance of a HygroClip2 within its stated specification is based on temperature

adjustment being made at 23°C. AutoCal does permit the adjustment to be made at other temperatures.

However, this is only recommended when the operational environment of the probe being adjusted is

only ever at another specific temperature, e.g. a stability cabinet at 40°C. Adjustments made at

temperatures other than 23°C risk comprising the probe’s ability to meet its specification within the

range of optimal performance: 18 to 28°C.

Please note that HW4 cannot be used at the same time as AutoCal.