3.4 HygroGen2 Touchscreen Interface

The HygroGen2 controller consists of four top level screens that are accessed using the forward or back

buttons in a loop:

Please note: the Programmer Screen is replaced by the AutoCal Screen when AutoCal is activated. The

standard Programmer functionality is still available, however. Please see

section 4.1

for more details.

3.4.1 Controller Home Screen

1) Menu (drop down)

2) Chamber temperature

3) Temperature set-point

4) Heating power

5) Temperature control

6) Chamber humidity

7) Humidity set-point

8) Humidification power

9) Humidity control

10) Previous screen

11) Prog/ACal function

12) HW4 software

13) Power off

14) Next screen

Navigating to Other Screens (items 1, 10 & 14)

The user can navigate to different screens using the drop-down menu (item 1) or can scroll through the

four main screens using items 10 & 11.

Setting Temperature and Humidity Set-Points (items 3 & 7)

At any time, the user can change humidity or temperature set-points by simply touching the set-point

on screen. A pop-up will appear and either the touch screen or, when connected, the keyboard/mouse

may be used to input desired values.

Note: To aid a smooth and efficient transition between temperature set-points, the controller is

programmed to suspend humidification during changes in temperature. Once the chamber temperature

is within 0.7°C of the temperature set-point, the humidity control is activated.